Specifications
Synergy Nano Technical Manual, Revision -
Page 55
Output 18 Control Type
The Output 18 Control Type selects the type of control logic for Vent/Boost Cool output.
Vent / Boost Cool Logic Diagram
OT18 Selects the Control logic for the Vent – Boost
Cool device
L14 Time Delay Boost Cool Time delay (in seconds) required before Boost
Cool is enabled
Note: Turn On Timer is reset when Channel 1 PID Cool < 100%
Low Limit, Channel 1 and 2
This parameter sets the lowest value a user can enter for the channel, whether temperature, humidity or
pressure. This is not the alarm setting and does not affect alarm setpoint. See Section 7.2 Maintenance:
Alarm Functions for more alarm information.
High Limit Channel 1 and 2
This parameter sets the highest value a user can enter for the channel, whether temperature, humidity or
pressure. This is not the alarm setting and does not affect alarm setpoint. See the section on Alarms
in the
manual for more alarm information.
Temperature RTD Curve (removed starting Version 2.6.8)
Your Synergy Nano chamber accepts Platinum RTD (Resistance Temperature Detector) sensors.
Starting in Version 2.6.8, the RTD Temperature Curve setting is determined from the
SETUP/Calibration/Input/RTD screens separately for each sensor.
Vaisala Temperature Compensation (removed starting Version 2.6.8)
Your Synergy Nano chamber accepts Vaisala Relative Humidity sensors in temperature compensated and
uncompensated versions.
Starting in Version 2.6.8, the Vaisala compensation is made from the SETUP/Calibration/Input/Analog Input
Screens.