Specifications
Synergy Nano Technical Manual, Revision -
Page 77
6.8 LCD Settings
LCD Brightness adjust is not available on the Synergy Nano
6.9 Chamber Setup
The Chamber Setup Directory is used for factory setup. The chamber type specified in the Synergy Nano
must match the chamber that it is controlling. The Chamber Type setting maps software outputs to chamber
hardware. Each chamber type has as specific map that is unique to that type of chamber. The operator
should NEVER change this setting. It should only be modified by a qualified technician.
CHAMBER SETUP \ SETUP DIRECTORY SYNERGY NANO
Chamber Type Screen
(Screen Editor)
Nano Temperature Only
Nano Temperature/Temperature *
Nano Temperature/Humidity *
Nano Temperature/Pressure *
Nano Temperature/Vibration
CHAMBER SETUP \ SETUP DIRECTORY SYNERGY NANO
Chamber Type Screen
(Screen Editor)
Generic Temperature Only
Generic Temperature/Temperature *
Generic Temperature/Humidity *
Generic Temperature/Humidity Single Stage
Generic Temperature/Pressure *
Generic Temperature/Humidity/Pressure
Generic Temperature/Vibration
Retro Temperature Only
The Chambers Types in the list above that are identified with * support VersaTenn retrofit fit applications.
WARNING: Incorrect Chamber Type settings may cause severe damage to your chamber.