Specifications
170
Turbo-V 701 Navigator
130.8 (5.1)
229.1(9.0)
314.8 (12.4)
A
196.1 (7.7)
253.2 (10.0)
(WATER COOLING)
(WATER COOLING)
CFF 10"
VIEW FROM "B"
109.7 (4.3)
Ø224.0 (8.8)
Ø240.0 (9.5)
324.9 (12.8)
146.5 (5.8)
126.0 (5.0)
46.0 (1.8)
70.0 (2.8)
200.6 (7.9)
115.1 (4.5)
ISO 200
B
Ø203.7 (8.0)
VENT
PORT
ISO 200-F
12°0'
=
=
157.1 (6.2)
VIEW FROM "A"
166.0 (6.5)
FORELINE
KF 25
324.7 (12.8)
PURGE
PORT
ELBOW 90°
ELECT.
CONNECTOR
WATER
COOLING 1/8 G
WATER
COOLING 1/8 G
Ø285.0 (11.2)
Ø203.7 (8.0)
Outline Drawings
Dimensions: millimeters (inches)
Technical Specifications
Pumping speed (with inlet screen)
CF 10” or ISO 200: N
2
: 690 l/s He: 620 l/s H
2
: 510 l/s
Compression ratio N
2
: 1 x 10
9
He: 1 x 10
7
H
2
: 1 x 10
6
Base pressure *(with minimum recommended forepump) <1 x 10
-10
mbar (< 1 x 10
-10
Torr)
Inlet flange CF 10” ISO 200 ISO 200-F bolted
Foreline flange KF 25
Rotational speed 42,000 rpm
Start-up time <5 minutes
Recommended forepump Mechanical: Varian DS 402
Dry scroll: Varian TS300
Operating position Any
Cooling requirements Natural air convection
Water optional (use water with electrical
conductivity ≤ 500 µS/cm)
Bakeout temperature 120 °C at inlet flange max. (CF flange)
80 °C at inlet flange max. (ISO flange)
Vibration level (displacement) <0.01 µm at inlet flange
Weight kg (lbs) ISO flange 19.4 (43.0)
CF flange 25.5 (54.2)
* According to PNEUROP 5608.
SEM version available on request