Specifications
172
Turbo-V 1001 Navigator
Ø285.0 (11.22) Ø335.0 (13.19)
Ø232.0 (9.13) Ø232.0 (9.13)
294.3 (11.48)
158.1 (6.17)
ISO 160
ISO 200
ISO 250
ISO 200-F ISO 250-F
158.1 (6.17)
252.8 (9.86)
279.4 (10.90)
52.0 (2.03)
CFF 10"
322.7 (12.58)
228.5 (8.91)
122.8 (4.79)
200.6 (7.82)
70.0 (2.73)
52.0 (2.03)
ELEC.
CONNECTOR
WATER
COOLING
WATER
COOLING
VENT
102°0'
WATER
COOLING
ELEC.
CONNECTOR
WATER
COOLING
VENT
102°0'
102°0'
323.5 (12.74)
350.6 (13.80)
Ø232.0 (9.13)
Outline Drawings
Dimensions: millimeters (inches)
Technical Specifications
Pumping speed (with inlet screen)
ISO 160: N
2
: 790 l/s He: 820 l/s H
2
: 860 l/s
CFF 10” or ISO 200: N
2
: 950 l/s He: 870 l/s H
2
: 900 l/s
ISO 250: N
2
: 1,050 l/s He: 900 l/s H
2
: 920 l/s
Compression ratio
ISO 160 N
2
:1 x 10
9
He: 1 x 10
7
H
2
:1 x 10
6
CF 10”/ISO 200 - ISO 250 N
2
:1 x 10
9
He: 1 x 10
7
H
2
:1 x 10
6
Base pressure* (with recommended forepump) <1 x 10
-10
mbar (< 1 x 10
-10
Torr)
Inlet flange ISO 160 ISO 200 ISO 200-F bolted CF 10” ISO 250 ISO 250-F bolted
Foreline flange KF 40
Rotational speed 38,000 rpm
Start-up time <4 minutes
Recommended forepump Mechanical: Varian DS 402
Dry scroll: Varian TS300
Operating position Any
Cooling requirements Forced air or water
Bakeout temperature 120 °C at inlet flange max. (CF flange)
80 °C at inlet flange max. (ISO flange)
Vibration level (displacement) < 0.01 µm at inlet flange
Weight kg (lbs)
ISO 160 flange 19 (41.8)
ISO 200 flange 19.4 (43.0)
CF 10” flange 25.5 (54.2)
ISO 250 flange 21.2 (46.6)
* According to PNEUROP 5608.
SEM version available on request