Technical information

SimradEK60
Note
Alltransducersmustbehandledasdelicateitems.Anywrongfulhandlingmaydamage
thetransducerbeyondrepair .
Donotactivatethetransducerwhenitisoutofthewater .
Donotliftthetransducerbythecable.
Donotsteponthetransducercable.
Donothandlethetransducerroughly,avoidimpacts.
Donotexposethetransducertodirectsunlightorexcessiveheat.
Donotusehighpressurewater ,sandblastingormetaltoolstocleanthetransducerface.
Donotusestrongsolventstocleanthetransducerface.
Calibrationprocedures
Thissectionholdsthecalibrationprocedures.
Topics
Checkinstallationonpage38
Anchoringonpage40
Winchesonpage41
Attachingthesphereonpage42
Loweringthesphereonpage43
Referencetargetonpage43
Calibrationonpage44
Viewsonpage46
Dataeditingonpage48
Updatingtransducerparametersonpage49
Previouslyrecordeddataonpage49
Ofinecalibrationonpage49
Checkinstallation
1Checkthattheechosounderandallthetransceiversandtransducersareinstalled
correctly,andthattheyareallfullyfunctional.
2Measurethewatersalinityandtemperaturebetweenthetransducerandthesphere
depth.
3Calculatetheaveragesalinityandtemperaturevalues,andenterthesedatainthe
Environmentdialog.
Environmentonpage110
38
164692/D