Technical information
SimradEK60
4Power
Thisparameterdisplaysandcontrolsthetransmitter’soutputpower.Outputpower
islimitedeithertothemaximumratingofthetransducer,orthemaximumratingof
thetransmitter,whicheveristhesmallest
.
5Depth
Thisisthedepthofthetransducerfacerelativetothewatersurface.
Inordertomeasurecorrectwaterdepth,theEK60needstoknowthevertical
distancebetweenthevessel’swaterlineandtheacousticfaceofeachtransducer.
Thedepthofeachindividualtransducermustbedenedmanually.Enterthedepth
asapositivenumber.
Example2Transducerdepth
Ifthedistancefromthewaterlineanddowntothetransducerfaceis3,5
meters,enter+3,5m.
Multiplexeroperation
WhentheEK60Scienticechosounderisoperatedwithamultiplexer,bothtransceiver
channelsareshownintheNormalOperationdialog.
Inthismode,youarepermittedtochangetransducerdepth,pulseduration,sample
intervalandbandwidthforthesecondtransducer,butyoucannotchangedthemode
andpoweroutput.
Relatedtopics
•Operationmenu;functionsanddialogsonpage75
•Outputpoweronpage231
•Pulsedurationonpage233
Replay
TheReplaydialogisopenedfromtheOperationmenu.
Purpose
ThepurposeoftheReplaydialogistochoosethele(s)tobeusedforplayback.The
lenameslistedhavebeengeneratedautomaticallyduringrecording,andeachleis
identiedwiththetimeanddateitwasmade.