- 1 Dear Customer,
- 2 General data
- 3 General description
- 4 Safety
- 5 Installation and startup
- 5.1 Transport and unpacking
- 5.2 Disposal of packaging materials
- 5.3 Installation site
- 5.4 Commissioning
- 5.4.1 Functional elements
- 5.4.2 Supplied tools
- 5.4.3 Touch display description
- 5.4.4 Illumination of the processing chamber, LED light strip, and On/Off button
- 5.4.5 Using the processing chamber screen
- 5.4.6 Installation
- 5.4.6.1 Connecting to the PC via LAN
- 5.4.6.2 Connecting the unit to the power supply
- 5.4.6.3 Installing the unit
- 5.4.6.4 Connecting the suction device (optional)
- 5.4.6.5 Connecting to the PC via WLAN (option)
- 5.4.6.6 Operating several production units at one access point
- 5.4.6.7 Connecting to the PC via the wireless interface (optional)
- 5.4.7 Filling the water tank
- 5.4.8 Switching the unit ON and OFF
- 5.5 Repacking
- 5.6 Scope of supply
- 5.7 Storage
- 6 Operation
- 7 Service
- 7.1 Using the cleaning hose and the wet cleaning process
- 7.2 Changing filter bags and HEPA filters
- 7.3 Changing the water
- 7.4 Instruments
- 7.5 Care, cleaning agents, and disinfectants
- 7.6 Cleaning surfaces
- 7.7 Replacing the main fuses
- 7.8 Changing the filter
- 7.9 Removing water from the unit
- 7.10 Using the tank cap opener
- 8 Technical description
- 9 Disposal
- 10 Consumable
- Index
4 Safety
4.5 Disturbance of data transmission
Dentsply Sirona
Operating Instructions CEREC Primemill
16
67 19 681 D3439
D3439.201.01.01.0209.2019
4.5
Disturbance of data transmission
Note on wireless communication
Data communication between the acquisition unit and the CEREC
Primemill production unit should preferably be established via the
wireless interface CEREC Radio Device or WLAN.
As for all wireless connections (e.g. cell phones), heavy utilization of the
available radio channels or shielding caused by building installations
(e.g. metal-shielded X-ray enclosures) may impair the quality of the
connection. This may become noticeable through a reduction in range
and/or a slower data transmission rate. In extreme cases, it will be
impossible to establish a wireless connection at all.
Dentsply Sirona has selected the best possible configuration for data
communication via the wireless interface (CEREC Radio Device) or
WLAN, which generally ensures perfect functioning of this connection.
However, in individual cases unrestricted wireless data communication
may be impossible for the reasons mentioned above and/or due to local
circumstances. In such cases, a cable LAN connection should be
selected to ensure uninterrupted operation. If the only LAN interface on
the rear of the acquisition unit is occupied by another plug, remove this
wireless interface connection and instead connect the LAN cable with
the CEREC Primemill production unit.
4.6
Ventilation slots
Under no circumstances may the ventilation slots on the unit be
covered, since otherwise the air circulation will be obstructed. This can
cause the unit to overheat.
Do not spray liquids such as disinfectants into the ventilation slots. This
may lead to malfunctions. Use wipe disinfection only in the vicinity of the
ventilation slots.