5 Installation and startup
5.4 Commissioning
Dentsply Sirona
Operating Instructions CEREC Primemill
30
67 19 681 D3439
D3439.201.01.01.0209.2019
5.4.6.4.4 Automatic mode
ü The interface cable is connected.
Set the on/off switch to the position Auto.
Note on the automatic mode:
The production unit monitors the correct connection (interface cable and
suction tube) and the operation of the suction system during the running
processes.
5.4.6.4.5 Setting the suction power
Use the control dial (
A
) to set the suction power.
Recommendation
: Set the suction power to the minimum (min).
NOTE
We recommended replacing the filter bag for the CEREC suction after
approx. 120 restoration units. This number may differ depending on
the amount of zirconium oxide materials milled and suctioned.