Operating Instructions
Table Of Contents
- Table of contents
- 1 Dear Customer,
- 2 General data
- 3 General description
- 4 Safety
- 5 Installation and startup
- 5.1 Transport and unpacking
- 5.2 Disposal of packaging materials
- 5.3 Installation site
- 5.4 Commissioning
- 5.4.1 Functional elements
- 5.4.2 Supplied tools
- 5.4.3 Touch display description
- 5.4.4 Illumination of the processing chamber, LED light strip, and On/Off button
- 5.4.5 Using the processing chamber screen
- 5.4.6 Installation
- 5.4.6.1 Connecting to the PC via LAN
- 5.4.6.2 Connecting the unit to the power supply
- 5.4.6.3 Installing the unit
- 5.4.6.4 Connecting the suction device (optional)
- 5.4.6.5 Connecting to the PC via WLAN (option)
- 5.4.6.6 Operating several production units at one access point
- 5.4.6.7 Connecting to the PC via the wireless interface (optional)
- 5.4.7 Filling the water tank
- 5.4.8 Switching the unit ON and OFF
- 5.5 Repacking
- 5.6 Scope of supply
- 5.7 Storage
- 6 Operation
- 7 Service
- 7.1 Using the cleaning hose and the wet cleaning process
- 7.2 Changing filter bags and HEPA filters
- 7.3 Changing the water
- 7.4 Instruments
- 7.5 Care, cleaning agents, and disinfectants
- 7.6 Cleaning surfaces
- 7.7 Replacing the main fuses
- 7.8 Changing the filter
- 7.9 Removing water from the unit
- 7.10 Using the tank cap opener
- 8 Technical description
- 9 Disposal
- 10 Consumable
- Index
Dentsply Sirona
Operating Instructions CEREC Primemill
6 Operation
6.3 Machining process
67 19 681 D3439
D3439.201.01.01.0209.2019
43
6.3.4 Ending the machining processes
1.
When the machining process has been completed, open the
processing chamber door.
2.
Remove the restoration.
WARNING
Risk of injury on the remainder of the ceramic block
The remaining portion of the ceramic block may have sharp edges
(e.g.
A
) that could injure you if it is not removed carefully.
Always grasp the remainder of the ceramic block by its metal holder.
3.
Loosen the ball pressure screw.
4.
Remove the remainder of the ceramic block. When removing the
remaining block from blocks with 6mm diameter block holders,
make sure that the adapter sleeve remains in the machine.
5.
Close the processing chamber door.
CAUTION
Do not use defective milling and grinding results!
Machining results must be judged by the user (dentist or dental
technician) and must not be used if defects are detected!
NOTE
If you have not used the production unit for an extended period, we
recommend switching it off and then opening the processing chamber
door so that the processing chamber can dry out.