Operating Instructions
Table Of Contents
- Table of contents
- 1 Dear Customer,
- 2 General data
- 3 General description
- 4 Safety
- 5 Installation and startup
- 5.1 Transport and unpacking
- 5.2 Disposal of packaging materials
- 5.3 Installation site
- 5.4 Commissioning
- 5.4.1 Functional elements
- 5.4.2 Supplied tools
- 5.4.3 Touch display description
- 5.4.4 Illumination of the processing chamber, LED light strip, and On/Off button
- 5.4.5 Using the processing chamber screen
- 5.4.6 Installation
- 5.4.6.1 Connecting to the PC via LAN
- 5.4.6.2 Connecting the unit to the power supply
- 5.4.6.3 Installing the unit
- 5.4.6.4 Connecting the suction device (optional)
- 5.4.6.5 Connecting to the PC via WLAN (option)
- 5.4.6.6 Operating several production units at one access point
- 5.4.6.7 Connecting to the PC via the wireless interface (optional)
- 5.4.7 Filling the water tank
- 5.4.8 Switching the unit ON and OFF
- 5.5 Repacking
- 5.6 Scope of supply
- 5.7 Storage
- 6 Operation
- 7 Service
- 7.1 Using the cleaning hose and the wet cleaning process
- 7.2 Changing filter bags and HEPA filters
- 7.3 Changing the water
- 7.4 Instruments
- 7.5 Care, cleaning agents, and disinfectants
- 7.6 Cleaning surfaces
- 7.7 Replacing the main fuses
- 7.8 Changing the filter
- 7.9 Removing water from the unit
- 7.10 Using the tank cap opener
- 8 Technical description
- 9 Disposal
- 10 Consumable
- Index
Dentsply Sirona
Operating Instructions CEREC Primemill
7 Service
7.3 Changing the water
67 19 681 D3439
D3439.201.01.01.0209.2019
51
7.3
Changing the water
7.3.1 General information
NOTE
Coolant
Use distilled or demineralized water.
If a water change is due, a notification appears on your touch display.
Preventing odors
All coolant additives contain a biodegradable preservative. Despite this,
however, odors may still develop under unfavorable conditions.
Observe the following:
● Change the water at least once a week.
● With ambient temperatures above 25°C (77°F), change the water
every 2 to 3 days to prevent foul odors.
● Drain the tank if you do not intend to operate the unit for more than
one week.
● Clean the tank if the odors recur.
● Add the coolant additive DENTATEC and fill the tank up to the brim
with water. Let it stand for at least 24 hours and then rinse it out
thoroughly with water once again.
NOTE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates
plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
NOTE
Approved coolant additive
Use only DENTATEC as a coolant additive.