Operating Instructions
Table Of Contents
- Table of contents
- 1 Dear Customer,
- 2 General data
- 3 General description
- 4 Safety
- 5 Installation and startup
- 5.1 Transport and unpacking
- 5.2 Disposal of packaging materials
- 5.3 Installation site
- 5.4 Commissioning
- 5.4.1 Functional elements
- 5.4.2 Supplied tools
- 5.4.3 Touch display description
- 5.4.4 Illumination of the processing chamber, LED light strip, and On/Off button
- 5.4.5 Using the processing chamber screen
- 5.4.6 Installation
- 5.4.6.1 Connecting to the PC via LAN
- 5.4.6.2 Connecting the unit to the power supply
- 5.4.6.3 Installing the unit
- 5.4.6.4 Connecting the suction device (optional)
- 5.4.6.5 Connecting to the PC via WLAN (option)
- 5.4.6.6 Operating several production units at one access point
- 5.4.6.7 Connecting to the PC via the wireless interface (optional)
- 5.4.7 Filling the water tank
- 5.4.8 Switching the unit ON and OFF
- 5.5 Repacking
- 5.6 Scope of supply
- 5.7 Storage
- 6 Operation
- 7 Service
- 7.1 Using the cleaning hose and the wet cleaning process
- 7.2 Changing filter bags and HEPA filters
- 7.3 Changing the water
- 7.4 Instruments
- 7.5 Care, cleaning agents, and disinfectants
- 7.6 Cleaning surfaces
- 7.7 Replacing the main fuses
- 7.8 Changing the filter
- 7.9 Removing water from the unit
- 7.10 Using the tank cap opener
- 8 Technical description
- 9 Disposal
- 10 Consumable
- Index
Dentsply Sirona
Operating Instructions CEREC Primemill
7 Service
7.4 Instruments
67 19 681 D3439
D3439.201.01.01.0209.2019
55
7.
If an application is selected requiring the replacement of more than
one instrument, repeat the aforementioned process for the
remaining instruments.
NOTE
Cleaning cooling water nozzles
The cooling water nozzles in the processing chamber must be free
from limescale, grinding dust/milling dust deposits. The corresponding
cooling water jet must always strike the instrument accurately!
ü The cooling water nozzles are dirty.
➢ Clean the nozzles with the supplied tool.
NOTE
Use only suitable instruments!
Do not use any instruments of the devices CEREC MC XL / MC X /
MC / inLab MC XL or inLab MC X5.
Changing a defective instrument
If an instrument breaks during the processing process or an instrument
with low remaining service life is indicated during routine maintenance,
the instrument in the touch display is displayed in red. The touch display
also offers the option of replacing the defective instrument.
ü A dialog box is displayed on the touchscreen which specifies the
instrument to be changed or the application to be used.
1.
Select the instrument.
Ä The motors move to the position for changing the instrument.
Ä The dialog box for changing the instruments opens.
2.
Open the processing chamber door.
3.
Loosen the instrument with the torque wrench and pull it out
manually.
4.
Hold the colored rear end of the instrument approx. 1–3mm in front
of the RFID scanner.
Ä If the correct unit type is selected and read by the RFID
scanner, the touch display shows this. The motor to which the
instrument is to be integrated is also shown.
Ä If an incorrect or defective / used unit type is selected and read
by the RFID scanner, the touch display shows that another
(correct or new) instrument should be selected.
5.
Insert the instrument into the corresponding motor as shown on the
touch display. Tighten the corresponding chuck with the torque
wrench until a clicking sound can be heard.
6.
Close the processing chamber door.