5 Installation and startup
5.4 Commissioning
Dentsply Sirona
Operating Instructions CEREC Primemill
30
67 19 756 D3439
D3439.201.01.01.0209.2019
5.4.6.3.2 Process steps that have to be performed on the acquisition unit or
the PC
Automatic unit search
ü The unit is connected to the PC via a LAN cable or via WLAN.
1.
Click the
"Configuration"
button in the system menu.
2.
Click on the
"Devices"
button.
3.
Click on the
"Scan for New Devices"
button.
Ä All units connected to the PC are recognized. In the case of
new units, you will be prompted to enter a name.
4.
Enter a name and a short designation for the new unit.
Ä The unit name, the designation, and the software version are
then shown on the touch display.
Manual unit search
ü The unit is connected to the PC via a LAN cable or via WLAN.
1.
Click the
"Configuration"
button in the system menu.
2.
Click on the
"Devices"
button.
3.
Click on the
"Add Device (Manual)"
button.
4.
Set the network.
5.
Enter the network address which appears on the
"IP address:"
display once the production unit has been switched on.
6.
Click on the
"Ok"
button.
Ä The software attempts to contact the device.
If the connection fails, check the connection. If necessary, ask a
qualified technician.
Remove the unit
ü If you no longer require a unit (e.g. a unit is replaced), you can
remove it.
ü The unit is not in operation.
1.
Click the
"Configuration"
button in the system menu.
2.
Click on the
"Devices"
button.
3.
Click on the unit that you wish to uninstall.
4.
Click on the
"Delete Device"
button.
Ä You will be asked if you would like to remove the unit.
5.
Click on the
"YES"
button.
Ä
The device is removed.