Operating Instructions
Table Of Contents
- Table of contents
- 1 Dear Customer,
- 2 General data
- 3 General description
- 4 Safety
- 5 Installation and startup
- 5.1 Transport and unpacking
- 5.2 Disposal of packaging materials
- 5.3 Installation site
- 5.4 Commissioning
- 5.4.1 Functional elements
- 5.4.2 Supplied tools
- 5.4.3 Touch display description
- 5.4.4 Illumination of the processing chamber, LED light strip, and On/Off button
- 5.4.5 Using the processing chamber screen
- 5.4.6 Installation
- 5.4.6.1 Connecting to the PC via LAN
- 5.4.6.2 Connecting the unit to the power supply
- 5.4.6.3 Installing the unit
- 5.4.6.4 Connecting the suction device (optional)
- 5.4.6.5 Connecting to the PC via WLAN (option)
- 5.4.6.6 Operating several production units at one access point
- 5.4.6.7 Connecting to the PC via the wireless interface (optional)
- 5.4.7 Filling the water tank
- 5.4.8 Switching the unit ON and OFF
- 5.5 Repacking
- 5.6 Scope of supply
- 5.7 Storage
- 6 Operation
- 7 Service
- 7.1 Using the cleaning hose and the wet cleaning process
- 7.2 Changing filter bags and HEPA filters
- 7.3 Changing the water
- 7.4 Instruments
- 7.5 Care, cleaning agents, and disinfectants
- 7.6 Cleaning surfaces
- 7.7 Replacing the main fuses
- 7.8 Changing the filter
- 7.9 Removing water from the unit
- 7.10 Using the tank cap opener
- 8 Technical description
- 9 Disposal
- 10 Consumable
- Index
6 Operation
6.2 Calibrating the unit
Dentsply Sirona
Operating Instructions CEREC Primemill
42
67 19 756 D3439
D3439.201.01.01.0209.2019
Calibration procedure
1.
Select the symbol shown on the left in the touch display.
2.
Select
"Calibration"
and then
"Start"
.
3.
Register the calibration phantom on the machine (scanning the
code) and insert it in the machine.
4.
Close the door.
5.
Register the calibration pins (set 1) on the machine (RFID) and
insert them in the machine.
6.
Close the door.
Ä The calibration for set 1 is performed. The automatic calibration
starts and lasts approx. 4minutes. Wait until the calibration has
been completed.
7.
Remove the calibration pins from the machine.
8.
Close the door.
9.
Register the calibration pins (set 2) on the machine (RFID) and
insert them in the machine.
10.
Close the door.
Ä The calibration for set 2 is performed. The automatic calibration
starts and lasts approx. 4minutes. Wait until the calibration has
been completed.
11.
Remove the calibration pins from the machine.
12.
Close the door.
13.
Remove the calibration phantom.
14.
Close the door.
Ä The calibration is completed.