Operating Instructions

Table Of Contents
6 Operation
6.4 Manual block clamp
Dentsply Sirona
Operating Instructions CEREC Primemill
48
67 19 756 D3439
D3439.201.01.01.0209.2019
Ä The block is pressed laterally against the contact surface of the
block fixing and simultaneously pulled in axially. The plate of
the block holder thus rests on the block fixing.
Removing the adapter sleeve
1.
Loosen the ball pressure screw.
2.
Place the adapter sleeve removal tool in the inner groove (
D
) and
pull out the adapter sleeve.
Blocks with 10 mm diameter block holders
NOTE
Fasten the block tightly
If the block is not sufficiently tightened, it can lead to unsuitable
results and ceramic breakages.
Securely tighten the block with the block clamping tool with the
torque wrench, until you can hear a clicking sound.
Check to make sure that the block is seated very firmly.
1.
Place the block (
A
) directly into the block fixing.
2. Clamp
the ceramic block
securely
with the ball pressure screw (
B
).
Use the block clamp tool with torque wrench for this purpose.
Ä The block is pressed laterally against the contact surface of the
block fixing and simultaneously pulled in axially. The plate of
the block holder thus rests on the block fixing.