28.6.3 Installation
Tool: Breathing filter: Skan part ID 1288920
Tool for turning the filter cover
1.
Place the WirelessGT on a horizontal surface with the filter opening upwards.
2.
Disinfect the conical contact surface between the filter and device.
3.
Replace the removed breathing filter (3) with a new one of the same type. Touch the new filter only on the
outside!
2
1
3
4.
For detailed instruction please contact Skan service hotline!
5.
Close the filter cover (2) of the WirelessGT by means of the delivered tool (1).