Operating Manual
Table Of Contents
- Table of contents
- 1 Service and information address
- 2 Information for your safety
- 2.1 Residual risks
- 2.2 Minimum qualifications for machine operators, maintenance and service personnel
- 2.3 Regular maintenance of the device
- 2.4 Operating and maintenance work on device components of other manufacturers
- 2.5 Troubleshooting
- 2.6 Maintenance work on live device components
- 2.7 System modifications, add-ons and conversions
- 2.8 Protecting against electrostatic discharge
- 2.9 Exchanging device components
- 2.10 Revision mode / Revision operation (manual modification of function parameters)
- 2.11 Using cleaning and disinfection agents
- 2.12 Disposal and recycling
- 3 Device conformity
- 4 Scope of delivery
- 5 Transport and installation information
- 6 Device identification
- 7 Technical data
- 8 Information on device use
- 9 Layout
- 10 System overview
- 11 Operational concept
- 12 Commissioning
- 13 Battery pack charger / AC/DC adapter
- 14 Trolley
- 15 Network extender
- 16 Access control and user administration
- 17 Data display
- 18 Database for test parameters
- 19 Settings and configuration
- 20 Data storage
- 21 Start-up
- 22 Glove test using the WirelessGT
- 22.1 Preparation
- 22.2 Login
- 22.3 Test with "Auto start"
- 22.4 Test without "Auto start"
- 22.5 Stop test manually
- 22.6 Retest glove port
- 22.7 Finish test on the current glove port
- 22.8 Test list
- 22.9 Glove test in campaign
- 22.10 End of tests
- 22.11 Negative pressure device (option)
- 22.12 Print files
- 23 Data recovery
- 24 Exemplary process parameters
- 25 Messages
- 26 Cleaning and disinfection agents
- 27 Maintenance plan
- 28 Maintenance activities
- 29 Trouble shooting
- Index
Operating manual
WirelessGT
Glove tester
Document No._Version:
341174_D
CMS ID
14f8b65f920839adc0a8203200539573-13-en-GB
Print Date
29.04.2021
Document Status
released
Page No.
118 / 122
Revision Date
29.04.2021 / schwla1
28.6.3 Installation
Tool: Breathing filter: Skan part ID 1288920
Tool for turning the filter cover
1.
Place the WirelessGT on a horizontal surface with the filter opening upwards.
2.
Disinfect the conical contact surface between the filter and device.
3.
Replace the removed breathing filter (3) with a new one of the same type. Touch the new filter only on the
outside!
2
1
3
4.
For detailed instruction please contact Skan service hotline!
5.
Close the filter cover (2) of the WirelessGT by means of the delivered tool (1).