Operating Manual
Table Of Contents
- Table of contents
- 1 Service and information address
- 2 Information for your safety
- 2.1 Residual risks
- 2.2 Minimum qualifications for machine operators, maintenance and service personnel
- 2.3 Regular maintenance of the device
- 2.4 Operating and maintenance work on device components of other manufacturers
- 2.5 Troubleshooting
- 2.6 Maintenance work on live device components
- 2.7 System modifications, add-ons and conversions
- 2.8 Protecting against electrostatic discharge
- 2.9 Exchanging device components
- 2.10 Revision mode / Revision operation (manual modification of function parameters)
- 2.11 Using cleaning and disinfection agents
- 2.12 Disposal and recycling
- 3 Device conformity
- 4 Scope of delivery
- 5 Transport and installation information
- 6 Device identification
- 7 Technical data
- 8 Information on device use
- 9 Layout
- 10 System overview
- 11 Operational concept
- 12 Commissioning
- 13 Battery pack charger / AC/DC adapter
- 14 Trolley
- 15 Network extender
- 16 Access control and user administration
- 17 Data display
- 18 Database for test parameters
- 19 Settings and configuration
- 20 Data storage
- 21 Start-up
- 22 Glove test using the WirelessGT
- 22.1 Preparation
- 22.2 Login
- 22.3 Test with "Auto start"
- 22.4 Test without "Auto start"
- 22.5 Stop test manually
- 22.6 Retest glove port
- 22.7 Finish test on the current glove port
- 22.8 Test list
- 22.9 Glove test in campaign
- 22.10 End of tests
- 22.11 Negative pressure device (option)
- 22.12 Print files
- 23 Data recovery
- 24 Exemplary process parameters
- 25 Messages
- 26 Cleaning and disinfection agents
- 27 Maintenance plan
- 28 Maintenance activities
- 29 Trouble shooting
- Index
Operating manual
WirelessGT
Glove tester
Document No._Version:
341174_D
CMS ID
14f8b65f920839adc0a8203200539573-13-en-GB
Print Date
29.04.2021
Document Status
released
Page No.
71 / 122
Revision Date
29.04.2021 / schwla1
21.2.5.2 Process enabled
You can store max 9 processes. If you don't need a process for a short time, you can disabled it on this field to
avoid tests by mistake.
●
The option "Process enabled" is set in the "Process" screen.
You can use for tests only enabled processes. Oherwise an alarm message is dispayed - test failed.
21.2.5.3
Small volume process
To check the self-tightness of the WirelessGT, the option “small volume process” has to be enabled. This spe-
cial process allows to build up the pressure more slowly, in order not to overshoot too much.
✓
The option "Small volume process" is set in the "Process" screen.
Start testing.
🡆 The test flow is the same like by testing gloves.
21.2.6
Campaign mode
✓ The option "Enable campaign mode" is set in the "General configuration / Process" area of the "Setup"
screen.
1.
Click the [Setup] tab on the screen.
🡆
The "Setup" screen appears:
2.
Click in the next free "Campaign mode" input field and assign the new campaigne name.
3.
Save settings.
4.
Enter the campaign number into the "Campaign" input field of each RFID you have to test within the cam-
paign.
5.
Save RFID's.