21.2.5.2 Process enabled
You can store max 9 processes. If you don't need a process for a short time, you can disabled it on this field to
avoid tests by mistake.
The option "Process enabled" is set in the "Process" screen.
You can use for tests only enabled processes. Oherwise an alarm message is dispayed - test failed.
21.2.5.3
Small volume process
To check the self-tightness of the WirelessGT, the option “small volume process” has to be enabled. This spe-
cial process allows to build up the pressure more slowly, in order not to overshoot too much.
The option "Small volume process" is set in the "Process" screen.
Start testing.
🡆 The test flow is the same like by testing gloves.
21.2.6
Campaign mode
The option "Enable campaign mode" is set in the "General configuration / Process" area of the "Setup"
screen.
1.
Click the [Setup] tab on the screen.
🡆
The "Setup" screen appears:
2.
Click in the next free "Campaign mode" input field and assign the new campaigne name.
3.
Save settings.
4.
Enter the campaign number into the "Campaign" input field of each RFID you have to test within the cam-
paign.
5.
Save RFID's.