24.2 Two piece glove system
The parameters given should be viewed as proposed values. They are purely an example.
You will find explanations for the parameters used here: Chapter 11 ‘Operational concept’
on page 24
Two piece glove system
Glove type
Manufacturer: Piercan
Material: CSM
Sleeve type
Manufacturer: Piercan
Material: PVC
Adapter type
Manufacturer: Skan
Material: PP
Parameter
t
1
[sec] t
2
[sec] t
3
[sec] n p
1
[Pa] p
2
[Pa] p
3
[Pa]
150 1200 480 100 3000 50 145
1
Total test duration: app.31.5 [min]
1
Total test duration: time includs time to inflate test cover seal (t0), assumed with t(0)=60 sec and inflation time
(t1) assumed with t(1)=150 sec. This plus t(2) and t(3) results to t
total
= app.1 890 sec. = app.31.5 min.
24.3
Relaxation time
Glove material
Test pressure
History of the glove
Age of the glove
Too short waiting time / relaxation time has negative influence on the reproducibility of the
measurement!