Operating Manual

Table Of Contents
Operating manual
WirelessGT
Glove tester
Document No._Version:
341174_D
CMS ID
14f8b65f920839adc0a8203200539573-13-en-GB
Print Date
29.04.2021
Document Status
released
Page No.
97 / 122
Revision Date
29.04.2021 / schwla1
25.3 Alarm messages in the popup screen
25.3.1 Alarm list
Displayed
alarm mes-
sage
red LED
flashing
Meaning Measure
Radio Module fast
Establishing of
communication
to the coordi-
nator failed.
Test of ZigBee-communication by placement of device
near ZigBee coordinator (USB-Stick with antenna).
Pressure
sensor Glove
fast Sensor error
Check response in revision mode.
Replace pressure sensor glove.
Pressure
sensor Seal
fast Sensor error
Check response in revision mode.
Replace pressure sensor seal.
Fuse Blown fast
Blade fuse
inside device
blown (circuit
protection)
First of all identification of the cause of the over-cur-
rent. After rectification of the defect replacement with
same type and rating of fuse (this for continued protec-
tion against risk of fire).
Timeout seal
inflate
slow
The pneumatic
seal couldn’t be
inflated within
the specified
time.
Change / extend parameter [t0].
Second measure: Check integrity of the affected pneu-
matic seal.
Timeout glove
inflate
slow
The glove
couldn’t be
inflated within
the specified
time.
Change / extend parameter [t1].
Second measure: Check visually the integrity of the
affected glove.
Battery charge
too low
slow
Capacity of
device battery
too low!
Charge device battery.
Process param-
eters invalid/
disabled
slow
Stop of test
because trans-
ferring process
to end device
failed!
Check communication with end device located directly
near monitoring system (coordinator).
Default settings
loaded
slow self-explanatory na
Pulse pressure
too low
slow
Hold pressure
can’t be estab-
lished!
Check visually the integrity of the affected glove.
Seal pressure
too low
slow
Threshold limit
[p00] of the
seal reached
Check of integrity of seal respectively pneumatic sup-
port circuit.