Battery Maintenance, Continued
Recharging the
battery
Follow the steps in the table below to recharge the battery using the supplied
charger.
Note:
If the battery is low the sampling process will terminate and the low
battery screen will display.
Step Action
1 Plug the battery charger into an electrical outlet.
2 Plug the battery charger into the battery charge jack on the top
panel of EPAM-5000 Battery.
Results:
• The battery charge begins. LED indicator should be Red.
• Unit must be off or the battery must be removed from
instrument for LED to be Red and charge battery.
• When LED is green battery is fully charged.
Note:
Recharging time is approximately 22 hours when the
instrument is not operating, and approximately 24 hours while the
monitor is sampling.
CAUTION: Do not charge in a hazardous environment. Use only
the EDC approved charger designed for the EPAM-5000.
Continued on next page