Manual

Aluminum Cyclone
25 mm - Catalog No. 225-01-01
37 mm - Catalog No. 225-01-02
The SKC Aluminum Cyclone
The SKC Aluminum Cyclone is a lightweight respirable dust sampler that is used with a
lter loaded into a three-piece lter casse e. The lter material, pore size, and the support
pad must be selected as speci ed by the sampling method used. The cyclone separates the
dust particles according to size. The respirable particles collect on a lter for analysis while
larger particles fall into the grit pot and are discarded.
Flow rate: 2.5 L/min to conform to the ISO/CEN particle size selection criteria and
OSHA proposed rule for silica (4.0-μm 50% cut-point)*
2.8 L/min for alternate applications (3.5-μm 50% cut-point)
* As previously published, a ow rate of 2.6 L/min will give a 4 μm cut-point, however, 2.5 L/min
will give a be er match over the entire curve.
Determined using experimental data obtained at ows from 2.0 to 4.0 L/min
Note: The ACGIH/ISO/CEN curve requires using the ow rate that minimizes the bias of the
cyclone separation from the convention curve.
Operation
Assembly
1. Disassemble a three-piece casse e and set
aside inlet section (usually marked “inlet”).
Remove plug from casse e outlet.
2. Select a lter and support pad as speci ed
in the sampling method. Insert casse e ring
(middle) section into casse e outlet (see Figure 1).
3. Insert cyclone into casse e ring and press
rmly. Leave grit pot in place during calibration
and sampling.
4. Insert cyclone/casse e assembly through
large opening of lter casse e holder (see Figure
3). Ensure assembly is seated rmly in holder
by inserting the cyclone‘s side pin (small round
nodule on rim of cyclone) into the notch in the
lter casse e holder.
5. Secure assembly in holder by stretching the
spring-loaded hold-down plate over the casse e.
Insert adapter (located in end of rubber tubing)
into casse e outlet.
Calibration
1. Prepare a cyclone/ lter casse e assembly (see Assembly).
Calibrate the cyclone with grit pot in place.
2. Moisten O-ring inside large end of calibration chamber (Cat. No. 225-01-
03, see Figure 2) and push chamber over cyclone stem until it ts snugly.
3. Connect the Tygon
®
tubing a ached to the lter casse e holder to the inlet
of a sampling pump. Using exible tubing, connect the calibration chamber to
a calibrator.
4. Following instructions in the pump and calibrator manuals, calibrate to
required ow rate (see reverse side).
5. Disconnect calibrator and replace lter used to set ow with a fresh
weighed lter (unused) for sample collection.
Sampling
1. Ensure ow rate has been
calibrated (see Calibration).
Warning: Leave grit pot in place
during calibration and sampling.
2. Connect the Tygon tubing a ached
to the lter casse e holder to the inlet of
a sampling pump.
3. Clip cyclone/ lter casse e assembly
onto a worker‘s collar or pocket near the
breathing zone.
4. Start pump and record pertinent
details.
5. At the end of the sampling period,
stop pump and record nish time.
Remove cyclone/ lter casse e assembly from casse e holder. Separate cyclone from
casse e. Reassemble inlet section of casse e, ensuring the inlet and outlet are sealed with
plugs. Send casse e and all data to a laboratory for analysis.
Note: Particles collected in the grit pot do not represent any part of the respirable dust sample and
should be discarded.
6. Prior to further sampling, clean all parts of the cyclone, including the interior of the grit
pot, with mild soapy water. The cyclone may be wiped with a clean dust-free tissue, air
dried, blown dry, or wiped with isopropyl alcohol.
Accessories Catalog No.
Three-piece Filter Cassette Blanks 25 mm
37 mm
225-3-25LF
225-3LF
Filter Cassette Holder 25 or 37 mm 225-1
Calibration Chamber 25 or 37 mm 225-01-03
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides
SKC’s sole liability and the buyers exclusive remedy. To view the complete SKC Limited
Warranty and Return Policy, go to h p://www.skcinc.com/warranty.asp.
Form 3803 Rev 1402
Figure 1
Exploded view
of the cyclone/ lter
cassette assembly
Figure 3
Cyclone with lter cassette in holder
Tygon
Tubing
Rubber
Tubing
Cyclone/Filter
Cassette Assembly
Filter
Cassette
Holder
Grit Pot
Collar Clip
Adapter
Air Flow
Figure 2
Calibration
Chamber
For more information: See SKC Technical Note, Publication 1519, at www.skcinc.com.