Owner's manual

Preloaded Coated Filters
Cat. No. 225-9002
SKC coated lters are shipped preloaded in the casse es with end plugs and
shrink bands in place. No assembly is required; they are ready to use.
Method & Chemical: OSHA 42 (modi ed*) for 1,6-hexamethylene diisocyanate
(HDI), toluene-2,6 diisocyanate (2,6-TDI), and toluene 2,4-diisocyanate
(2,4-TDI). OSHA 47 (modi ed*) for methylene bisphenyl isocyanate (MDI).
ASTM D5836 (modi ed*) for 2,4-TDI and 2,6-TDI only
* OSHA Method 42 speci es a lter loading of 0.1 mg of 1-(2-pyridyl) piperazine. OSHA 47 and
ASTM D5836 specify a lter loading of 1 mg of 1-(2-Pyridyl) piperazine; however, Coated
Filter 225-9002 (loading of 2 mg to increase stability and shelf-life of lters) is acceptable for
both sampling methods.
Filter & Coating: Glass ber lter and cellulose support pad coated with
2.0 mg of 1-(2-pyridyl) piperazine loaded into a three-piece 37-mm casse e.
Use in the open-face sampling con guration.
Prior to Sampling: Store at 39.2 F (4 C) for up to 6 months.
Sample Stability: Submit extracted samples to a laboratory as soon as possible
a er sampling. Store at 39.2 F (4 C) for up to one week if delay is unavoidable.
For information on other available
coated lters, request SKC Publication 1219.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 Fax: 724-941-1369
www.skcinc.com
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC's
sole liability and the buyer's exclusive remedy. To view the complete SKC Limited Warranty and
Return Policy, go to http://www.skcinc.com/warranty.asp.
How to Use the SKC Preloaded Coated Filters
Do not expose coated lters to high temperatures or prolonged periods in
direct sunlight.
1. Select a coated lter casse e for calibrating pump ow rate. A red plug secures
the casse e inlet, which is clearly marked “inlet,” and a blue plug secures the outlet.
Set up the sampling train for calibration with the representative lter casse e in
line. For “open-face” sampling, remove the outlet plug and the casse e inlet section
then connect the casse e to the sampling train. For details on se ing up a sampling
train, refer to the SKC Sample Setup Guide “Sampling Train - Filters and Cyclones”
(SKC Publication 1166).
2. Calibrate the pump to the desired ow rate using a lm owmeter or calibrated
precision rotameter. For details on calibrating a pump, refer to the SKC Sample Setup
Guide “Calibrating a Pump Using a Film Flowmeter” (SKC Publication 1163).
3. Replace the casse e used for calibration with a fresh coated lter casse e
for sample collection. Attach the cassette to a workers collar, as close to the
breathing zone as possible, by using a filter cassette holder (SKC Cat. No.
225-1). Sample for the time speci ed in the method and record the time.
4. Remove the lter casse e at the end of the speci ed sampling period and
replace the inlet and both plugs. Recheck the ow rate using the same casse e
and owmeter used for calibration (in Step 2) to ensure that the ow rate has not
changed by more than 5%.
5. Don disposable gloves. Remove casse e inlet section and use clean forceps to
remove lter from casse e. Place lter in a glass vial containing eld extraction
solution following the procedure in Appendix B of OSHA Directive CPL 03-00-017
(National Emphasis Program on Isocyanates) at www.osha.gov/OshDoc/Directive_
pdf/CPL_03-00-017.pdf .
6. Submit extracted samples to a laboratory as soon as possible a er sampling.
Store at 39.2 F (4 C) for up to one week if delay is unavoidable. Appropriately
package and ship samples, blanks, and all pertinent data to a laboratory for analysis.
Refrigerated shipment is not required. Follow all applicable hazardous materials
shipping restrictions and requirements. See Appendix B of OSHA Directive CPL 03-
00-017 for details.
Form 40031 Rev 1403