User Manual

Preloaded Coated Filters
Cat. No. 225-9009
SKC coated lters are shipped preloaded in the cassettes with end plugs and
shrink bands in place. No assembly is required; they are ready to use.
Method & Chemical: OSHA 82 for acetic anhydride
Filter & Coating: 2 glass ber lters coated with 1-(2-pyridyl) piperazine.
Four-piece 37-mm cassette. Closed-face sampling con guration
Prior to Sampling: Store at 39.2 F (4 C) for up to 6 months.
Sample Stability: Store samples at 39.2 F (4 C) for up to 15 days.
For information on other available
coated lters, request SKC Publication 1219.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 Fax: 724-941-1369
www.skcinc.com
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC's sole liability and the buyer's exclusive remedy. To view the complete
SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.
How to Use the SKC Preloaded Coated Filters
1. Select one coated lter cassette for calibrating the ow rate. A red plug
secures the inlet, which is clearly marked “inlet,” and a blue plug secures the
outlet. Set up the sampling train for calibration with the representative lter
cassette in line. For “closed-face” sampling, remove the plugs and connect the
cassette to the sampling train. For “open-face” sampling, remove the outlet
plug and the cassette inlet section then connect the cassette to the sampling
train. For details on setting up a sampling train, refer to the SKC Sample Setup
Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired ow rate using a lm owmeter or
calibrated precision rotameter. For details on calibrating a pump, refer to the
SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC
Publication 1163).
3. Replace the cassette used for calibration with a fresh coated lter cassette
for sample collection. Attach the cassette to a worker’s collar, as close to the
breathing zone as possible, by using a lter cassette holder (SKC Cat. No.
225-1). Sample for the speci ed time interval and record the time.
4. Remove the lter cassette at the end of the speci ed sampling period and
replace both end plugs (and the inlet if necessary). Recheck the ow rate using
the same cassette and owmeter used for calibration (in Step 2) to ensure that
the ow rate has not changed by more than 5%.
5. Store samples at 39.2 F (4 C) for up to 15 days. Appropriately package
samples, blanks, and all pertinent data to send to a lab for analysis. Expedite
shipping of samples to laboratory. Friday shipments are not recommended.
Form 40038 Rev 1303

Summary of content (1 pages)