User guide

Coated Filters
Cat. No. 225-9011
Method & Chemical: MDHS 25/3 (U.K.) for Organic Isocyanates; PI, TDI,
HDI, MDI.
Filter & Coating: 25-mm glass ber lter coated with 1-(2-methoxyphenyl)
piperazine. Filters only, not loaded in casse es
Prior to Sampling: Store at 39.2 F (4 C) for up to 6 months.
Sample Stability: Store TDI at 39.2 F (4 C) for up to 90 days. No data is
available for PI, HDI, and MDI.
For information on other available
coated lters, request SKC Publication 1219.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 Fax: 724-941-1369
www.skcinc.com
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC's sole liability and the buyer's exclusive remedy. To view the complete
SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.
How to Use the SKC Coated Filters
1. Set up the sampling train for calibration with the representative lter
casse e in line. For “closed-face” sampling, remove the plugs and connect
the casse e to the sampling train. For “open-face” sampling, remove the outlet
plug and the casse e inlet section then connect the casse e to the sampling
train. For details on se ing up a sampling train, refer to the SKC Sample Setup
Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired ow rate using a lm owmeter or
calibrated precision rotameter. For details on calibrating a pump, refer to the
SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC
Publication 1163).
3. Replace the casse e used for calibration with a fresh lter casse e for
sample collection. Attach the cassette to a workers collar, as close to the
breathing zone as possible, by using a lter casse e holder (SKC Cat. No.
225-1). Sample for the speci ed time interval and record the time.
4. Remove the lter casse e at the end of the speci ed sampling period and
replace both end plugs (and the inlet if necessary). Recheck the ow rate using
the same casse e and owmeter used for calibration (in Step 2) to ensure that
the ow rate has not changed by more than 5%.
5. Store sample at 39.2 F (4 C) for up to 90 days. Appropriately package
samples, blanks, and all pertinent data to send to a lab for analysis. Expedite
shipping of samples to laboratory. Friday shipments are not recommended.