QuickTake 30 Owner's manual

10
To navigate presets and displays:
Press
or to scroll through presets or displays.
To select a run time preset and run the pump:
Scroll to desired run time preset.
Select preset by pressing [
]. The pump will start to run.
To run the pump:
Press [
].
To place the pump in hold:
Press [
].
To repeat a sample run:
From Done, reset (rSEt), or sampling error (SErr), press [
] to return to run
time preset. Press [
] to sample.
To turn off LED:
The pump automatically turns o LED a er it is idle for 4 minutes. User can
reactivate the LED by pressing and holding
.
To interrupt a run, terminate a sample, or reset a pump:
From Hold, press
. Sample reset (rSEt) will display. Press [] to reset the
pump or
to ignore the reset and return to Hold.
• To set ow or calibrate:
See Calibration and Se ing Flow for details.
1. Connect a calibrator to the pump inlet.
2. Scroll to any run time preset. Press and hold CAL for 2 seconds to enter
calibration mode (pump will start running).
3. Press or to change ow rate until desired ow appears on the
calibrator ( ow rate does not display on LED).
4. Press
and Stor will display. Press [] to save se ing or to ignore
changes.
Selecting the
Available Presets
Press or
Press
5
4
3
2
1
CONT
INTT
END
999
998
997
Press