QuickTake 30 Owner's manual

13
Tubing
Calibrator outlet Pump inlet
Spore trap cassette
Note: Use a second cassette adapter to
attach calibrator to a spore trap cassette.
Digital display:
Calibration mode
Calibration with a spore trap cassette
Cassette adapter
Digital display:
Store setting
Calibration and Setting Flow
Use a primary standard calibrator to calibrate pump ow rate.
Calibrate the ow rate with a representative sampling medium in line.
Before use, allow the pump to equilibrate a er moving it from one temperature
extreme to another.
1. Ensure the pump has run for 5 minutes before calibrating. Using 3/8-inch
ID exible tubing and appropriate adapters, connect the inlet of the pump
to the outlet of a representative sampling medium and a primary standard
calibrator to the inlet of the representative sampling medium (as shown
above).
2. Scroll to any run time preset. Press and hold the CAL
bu on for two seconds. The pump will start running
and a ashing CAL will appear on the LED.
3. Press or to increase or decrease the ow until
the desired ow rate is displayed on the calibrator.
4. Press
. A ashing Stor will appear on the LED.
To save new se ing: Press [
]. The pump
will stop running.
To ignore new se ing: Press
. The pump will
stop running.
Note: A security code is not needed to change ow
rate. It is only required when changing factory
se ings. To change factory se ings, see Advanced
Setup.