QuickTake 30 Owner's manual

5
Run Time Features: User-selectable features, user-adjustable presets (see
Advanced Setup)
Preset timed runs: 1, 2, 5, 10, or 15 min
Manually set continuous
run with manual shut-off: 1 to 999 min (repeats 1 to 999-min
runs inde nitely until user stops
pump or power supply is depleted)
Intermittent sampling: See Advanced Setup
Flow Fault Features: If the pump is unable to compensate due to excessive back
pressure, the pump will go into ow fault mode (see page 17 for
details). Flow fault features are enabled as the default setting.
Flow fault may be disabled at the user’s option.
See Advanced Setup.
Flow fault: Flow fault displays immediately
Pump shuts off after 5 seconds
Fault restart: Pump attempts restart every 10 seconds
up to ve times
Media Compatibility: Viable cascade impactors, spore trap cassettes (e.g., VersaTrap),
asbestos cassettes, microvacuum cassettes, and other impaction
samplers that require ows from 10 to 30 L/min
Tubing: Requires 3/8-in ID tubing
Housing: ABS plastic
Dimensions: 9.3 x 8.4 x 3.5 in (23.6 x 21.3 x 8.9 cm)
Weight: 4.8 lbs (2.2 kg)
Indicates a warning, caution, or note
www.skcinc.com