Manual

Soap Film Flowmeter
Cat. No. 303
SKC primary standard owmeters are designed, manufactured, and tested for accuracy; no calibration
is needed. Read the operating instructions thoroughly before operating the owmeter.
Principle of Operation
The air sampling pump to be calibrated is connected to and pulls air through the owme-
ters volumetric glass tube where a at soap bubble ( lm) is interposed into the ow path.
As the air ow causes the lm to move up the volume marks, travel time is measured using
a stopwatch. Flow rate can be calculated using the travel time and known tube volume. The
303 Portable Flowmeter Kit is calibrated to a primary standard ± 2% of the volumes marked
on the owmeter.
Assembly
SKC Model 303 is shipped ready to use. Film solution is supplied as well as a length of rubber
tubing to connect the owmeter to the pump.
Preparing for Operation
1. Remove the owmeter from the case, and place it upright on a level surface.
2. Remove the large rubber stopper from the top of the owmeter.
3. Tip the owmeter at an angle, and pour a small amount of lm solution into the top of
the glass tube. The liquid should run down into the rubber bulb at the bo om.
4. Continue adding lm solution until it completely lls the bulb and rises to just below
the side arm inlet. Squeeze the rubber bulb occasionally to release trapped air.
5. Replace the stopper on top of the owmeter.
Wetting the Walls of the Tube
The interior walls of the owmeter must be wet for the lm to travel without rupture. There
are two methods:
Method 1
1. Tip the owmeter to a horizontal position. Allow the lm solution to run out of the rub-
ber bulb and into the tube.
2. Maintaining the horizontal position, rotate the owmeter so that lm solution coats the
entire inside surface of the glass. Caution: Do not allow the solution to enter the metal
tube in the rubber stopper at the top of the owmeter.
3. Return the owmeter to an upright position and allow the liquid to ow back down
into the rubber bulb. Squeeze the bulb occasionally to release trapped air.
Method 2
1. Connect a pump to the owmeter (see Connecting a Pump).
2. Turn on the pump. Each successive lm that rises up the tube will wet the glass. The
walls are su ciently wet when the lm bubbles successfully reach the highest volume
mark on the owmeter.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 www.skcinc.com

Summary of content (4 pages)