Owner manual

Hydrogen Cyanide Passive Sampler
Cat. No. 590-400
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
Introduction
The SKC Hydrogen Cyanide (HCN) Passive Sampler, speci ed in OSHA Method
1015, provides accurate sampling of hydrogen cyanide from 0.44 to 20 ppm for
either 15-minute or 8-hour samples. The uptake rate for the HCN Passive Sampler
is 28.4 ml/min. The sampler design allows the user to load 600 mg of soda lime
sorbent (supplied in a vial) into the sampler housing in the eld immediately before
sampling and transfer the sorbent from sampler to vial in the eld immediately
a er sampling as speci ed in the method. Samples are desorbed with water and
analyzed by ion chromatography/electrochemical detector. The HCN Passive
Sampler is designed for single use.
Performance Pro le
Sorbent/Amount: Soda Lime, 600 mg (supplied in glass vial)
Housing Material: Nylon
Diameter: 1.4 in (3.5 cm)
Length
(including clip): 2.5 in (6.3 cm)
Measuring Range: 0.44 to 20 ppm
Analysis: Solvent extraction; ion chromatography (IC) with electro-
chemical detector (ELCM)
Sample Storage: 3 weeks at 68 F (20 C) in storage studies
Note: OSHA Method 1015 recommends that samples be sent
to a laboratory as soon as possible. If delay is unavoidable, store
samples in a refrigerator as a precaution.
Sample Time: Validated for 15 min and 8 hrs
Sampling Rate: 28.4 ml/min (4.40% RSD)