ULTRA III Passive Samplers with or without Blank,
Containing Charcoal Sorbent (Solvent Extraction)
Cat. No. 690-105, 690-105-NB, or 690-205
Analysis
1. a. For Cat. No. 690-105 or 690-105-NB (prefilled sampler):
Remove
the
sampler from
the
resealable pouch and proceed
to Step 2.
b. For Cat. No. 690-205 (sorbent in vial): Remove
the
lid from
the
vial and proceed to Step 4.
2. Lay the sampler on a at surface with the back of the sampler
facing upward.
3. With the clip oriented to the le , remove the plug from the
compartment containing the sample sorbent (on left side,
immediately above the di usion holes).
4. Hold the sampler or vial over a 3.7-ml glass vial. Use a larger
vial if using more than 2 ml of solvent.
5. Transfer the sorbent to a 3.7-ml or larger vial by tipping the
sampler or vial upside down over the vial. Gently tap to remove
any remaining sorbent.
6. Add typically 2 ml of desorption solvent to the sorbent in
the vial.
Add the solvent slowly.
7. Shake the sorbent for 30 minutes.
8. Analyze the sample by gas chromatography with the detector
speci ed in the method for the compound of interest.
9. Analyze the blank/correction sorbent, as needed, by repeating
Steps 2 through 8 for the sampler and Steps 1 and 4 through 8
for the sorbent vial.
Form 40127 Rev 1306
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyers exclusive remedy.
To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty.asp.
Analysis Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
Calculations
[(SW) - (BW)] (24.45 x 10
6
)
C =
(DE) (MW) (SR) (MIN) (PT)
Where:
C = Concentration of chemical (ppb)
SW = Sample weight by analysis (μg)
BW = Analyte weight in blank (μg)
PT = Pressure/temperature correction (see below)
DE = Desorption e ciency (see below)
MW = Molecular weight of chemical
SR = Sampling rate (ml/min)
MIN = Sampling time (minutes)
The equation above is correct for 25 C (298 K) and standard
atmospheric pressure (760 mm Hg). To convert to other temperatures
and pressures the correction factor is:
PT = (T
1
/T
2
)
1.5
(P
2
/P
1
)
Where:
T
1
= Sampling site temperature (in Kelvin)
T
2
= 298 K
P
1
= Sampling site pressure (in mm Hg)
P
2
= 760 mm Hg
Desorption e ciency should be determined and expressed as a
decimal (e.g., 98% = 0.98).
For sampling rates and desorption solvents, go to www.skcinc.
com/PassiveGuide/default.asp.