Owner's manual

Battery Pack for
AirChek 2000 Sample Pumps
Battery Replacement Notes:
To retain history, ensure the pump has gone to SLEEP after the last run. The battery
pack should not be removed while the pump is running.
Replacing the battery pack may erase scheduled runs programmed with DataTrac
2000 software. After replacing the battery pack, use the software to reprogram the
scheduled runs in the AirChek 2000 pump.
Replacing the battery pack will erase sampling time, delayed start, and other settings
entered using the pump keypad. Reset these parameters after replacing the battery
pack.
1. To remove a battery pack, release the battery pack by removing the two security
screws located on the bottom of the battery pack. Pull the battery pack away from
the pump body.
2. To install a battery pack, carefully
align the power connector on the
replacement battery pack with the
terminal on the bottom of the pump
base plate and push the battery pack
into place.
3. Replace and tighten the two security
screws removed in Step 1.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
www.skcinc.com
Power
connector
Ba ery
pack
Terminal
Pump base
plate
Form #40059 Rev 1008
Ba ery
charging
jack
Computer
interface
port
Ba ery a ached to pump Ba ery detached from pump
Cover
Ba ery
charging
jack
Covers
Cautions
• Use of a repaired or rebuilt battery pack voids any warranty and the
UL Listing for intrinsic safety.
• Do not charge or operate pump from charger in hazardous locations!
Use only an SKC-approved charger and battery pack designed for the
AirChek
®
2000 Sample Pump to ensure reliable performance. Failure to
do so voids intrinsic safety and any warranty.
• Failure to follow warnings and cautions voids any warranty.
4. Charge the new battery pack attached to or detached from the pump using only
the charging jack on the back of the battery pack.
www.skcinc.com
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides
SKC’s sole liability and the buyers exclusive remedy. To view the complete SKC Limited
Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.