Use Instructions

F401-01(A-190102) Sky Labs Inc.
RATED power frequency
magnetic fields
(50/60Hz)
IEC 61000-4-8
30A/m
30A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3V
0.15 MHz 80MHz
6V in ISM bands
between 0.15 MHz
and
80MHz
80% AM at 1kHz
3V
0.15 MHz 80MHz
6V in ISM bands
Between 0.15 MHz
and
80MHz
80%AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the EUT, including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Where P is the maximum output power
rating of the transmitter in watts(W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters(M).
Field strengths from fixed RF transmitters
as determined by an electromagnetic site
survey, should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol;
Radiated RF EM fields
IEC 61000-4-3
3V/m
80MHz 2,7GHz
80% AM at 1kHz
3V/m
80MHz 2,7GHz
80%AM at 1kHz