27
IEC 61000-4-5
neutral
that of a typical commercial or
hospital environment or typical
home environment
Voltage dips,
short
interruptions
and
voltage
variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5s
Not Applicable
Mains power quality should be
that of a typical commercial or
hospital environment or typical
home environment. If the user
of the iHelmet (Model:
LTD88Lite, LTD36Air,
LTD160Pro) requires continued
operation during power mains
interruptions, it is
recommended that the iHelmet
(Model: LTD88Lite, LTD36Air,
LTD160Pro) be powered from
an uninterruptible power supply
or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment or
typical home environment.
NOTE UT is the a.c. mains voltage prior to application of the test level
Guidance and manufacturer’s declaration – electromagnetic immunity
The iHelmet (Model: LTD88Lite, LTD36Air, LTD160Pro) is intended for use in the
electromagnetic environment specified below. The customer or the user of the iHelmet (Model:
LTD88Lite, LTD36Air, LTD160Pro) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted
RF
IEC
61000-4-6
3 Vrms
150 kHz to
80MHz
Not
Applicable
Portable and mobile RF communications
equipment should be used no closer to any
part of the iHelmet (Model: LTD88Lite,
LTD36Air, LTD160Pro), including cables,
than the recommended separation distance