Users Manual
iGlucose® 33
study results of typical iGlucose users
•
Like all commercially available blood glucose meters, your
iGlucose® Blood Glucose Meter result may vary slightly from
your actual blood glucose value. This may be due to slight
dierences in technique and the natural variation in the test
technology. The table below shows the results of a study
where 153 typical users used the iGlucose Meter to test their
blood glucose level. When glucose results were below 75 mg/
dL the iGlucose Meter gave results within 15 mg/dL of their
true blood glucose level 6 out of 6 times.
Dierence range between
the true blood glucose level
and the iGlucose Meter result
Dierence range between
the true blood glucose level
and the iGlucose Meter result
The percent (and number) of
meter results that match true
blood glucose level within x%
The percent (and number) of
meter results that match true
blood glucose level within x%
• When the users’ results were equal or greater than 75 mg/dL in
this study, the iGlucose Meter gave results within 15% of their true
blood glucose level 140 out of 147 times.
33.3%
2/6
43.5%
64/147
Within ±5
mg/dL
Within ±5
mg/dL
83.3%
5/6
81.6%
120/147
Within ±10
mg/dL
Within ±10
mg/dL
Within ±15
mg/dL
Within ±15
mg/dL
100%
6/6
95.2%
140/147