Users Manual
36 iGlucose®
If you are experiencing symptoms that are not consistent with your
blood glucose test results and you have made sure to follow all
instructions in this manual, contact your healthcare professional
immediately.
Consult your healthcare provider for appropriate disposal of used
test strips and lancets.
The iGlucose® Blood Glucose Meter displays results between 20 and
600mg/dL. If the test result is below 20 mg/dL, “LO” will appear on
the screen. Please repeat your test with by a new test strip.
If you still get a “LO” result, contact your healthcare professional.
If the test result is above 600 mg/dL, “HI” will appear on the screen.
Please repeat your test again with a new test strip.
If you still get a “HI” result, contact your healthcare professional.
about quality control testing
To ensure proper meter function, it is necessary to regularly perform
a quality control test.
Use Rightest® Control Solution GC550 when testing your