Users Manual

38 iGlucose®
required items for quality control tests
caution
To perform a quality control test, prepare the items below:
iGlucose® Blood Glucose Meter
iGlucose® Blood Glucose Test Strips
Rightest® Control Solution GC550
Each time you open a new bottle of control solution, write the
expiration date on the label. The Rightest® Control Solution
GC550 is good for 3 months after opening the bottle, or until
the expiration date printed on the label of the control solution,
whichever comes first.
Example
After use, wipe the top of the bottle cap with a clean wet tissue
and then with a dry one to remove any residue of the control
solution.
Close the bottle of control solution tightly immediately after using.
Keep control solution bottles out of reach of children.
WARNING
Do not apply blood when testing in the “Control Solution”
mode and when this icon is displayed. Testing blood in
this test mode will produce inaccurate blood results.