Users Manual

iGlucose® 43
caution
The control solution test should be conducted between
59-104°F (15-40°C) and in a location with a relative humidity
of 10-90%.
Do not touch the control solution to the sample port on the strip
before seeing the message “Apply a drop of control solution. The
iGlucose® Blood Glucose Meter performs an internal check that
takes 3 seconds. Touching the control solution to the sample port
before prompted will result in an error message.
Do not drip the control solution on to the sample port of the
test strip directly as this could flood and contaminate the meter
via the test strip port.
2
7.
6
C
Apply a drop of
control solution
Standby Mode