Users Manual
iGlucose® 45
If Rightest® Control Solution GC550 results are out of range, your
iGlucose® Blood Glucose Monitoring System may not be working
properly. Repeat the quality control test. If your control solution
results are still out of range, do not use the blood glucose meter
to test your blood glucose. Please contact the iGlucose Diabetes
Customer Support Center.
The iGlucose Blood Glucose Meter is able to automatically store a
maximum of 500 test results with time and date. If your meter has
stored 500 results, the newest test result will replace the oldest one.
The iGlucose Blood Glucose System maintains in the device memory
the averages of your blood glucose readings for the last 7, 14, 30,
60 and 90 days.
To obtain glucose readings averages:
• Press the “ ” button on the right side of the device.
The “Averages” menu appears.
•
Scroll to the desired option by using the “ ” or “ ” bu tton
then press the “Home” button to select:
°
Last 7 Days
°
Last 14 Days
°
Last 30 Days
°
Last 60 Days
°
Last 90 Days
Recalling Test Results
Obtaining Readings Averages