Users Manual Part 1

11
12. Insert the communication module into the pump
battery compartment. Hook the bottom of the
polemount adapter to the bottom of the pump,
and then snap the top of the polemount adapter
onto the base of the module.
NOTE: Align the teeth of
the polemount adapter
to the recessed opening
in the pump housing.
13. Use a CADD®-Solis Pump Key, standard
screwdriver, or a coin to tighten the screw to
attach the polemount adapter and secure the
communication module to the pump. Tighten the
screw until it is fully seated then another 1/8 turn
(45 degrees) approximately. Do not over tighten.
14. Before conguring the pump and rst use,
the communication module will need to be
fully charged. See the next section for more
information about charging the rechargeable
battery in the CADD®-Solis Communication
Module.
15. If the CADD®-Solis Communication Module is
being installed into a specic pump for the rst
time, the wireless settings for the pump will need
to be congured using the CADD®-Solis Network
Setup Utility software (available separately).
For more information, see the documentation
provided with the network setup utility software.
Charging the Rechargeable Battery
The communication module rechargeable battery is
charged whenever the CADD®-Solis pump is plugged
into AC power via the CADD®-Solis AC Adapter–even
when the pump is turned o.
Within the operating temperature range of 2°C to 40°C
(35°Fto 104°F), the rechargeable battery will become
fully charged in 10 hours or less.
Fully charge the rechargeable battery:
Before the first use
At least every 30 days while in regular use
At least once every 6 months while in storage
When the pump indicates the battery is low or
depleted
CAUTION: Failure to charge the rechargeable battery
during prolonged storage may cause deep battery
discharge, leading to premature battery replacement.
CAUTION: To avoid damaging the AC adapter
connector or the pump, do not use excessive force
or instruments such as pliers to attach or detach the
AC adapter connector.
Pump Recessed Opening
Polemount Adapter
Alignment
Teeth