Users Manual Part 1

17
Specications
Battery type
Rechargeable Lithium-ion battery (2 cells)
Equivalent Lithium content < 1.6 g
Battery capacity
18.7 Wh (when fully charged)
Battery output
3.6V (nominal)
Rechargeable battery weight
< 100 g
Operating environmental requirements
Temperature ............... 2°C to 40°C
(35°F to 104°F)
Humidity ................... 20% to 90% relative
humidity,
non-condensing
Atmospheric pressure ...... 70 kPa to 106 kPa
(10.2 psia to 15.4
psia)
Storage and transportation environmental
requirements
Temperature ............... –20°C to 60°C
(–4°F to 140°F)
Humidity ................... 20% to 90% relative
humidity,
non-condensing
Atmospheric pressure ...... 70 kPa to 106 kPa
(10.2 psia to 15.4
psia)
Product battery life
400 full cycles (charge / discharge) at room
temperature, 23°C (73°F)
Maximum charge time (fully depleted battery to
fully charged)
10 hours when plugged into AC Adapter
Pump and communication module run time (new
battery)
4 hours at pump nominal settings (see the
pump Operator's Manual for nominal settings
information)
Safety
The communication module complies with
Underwriters Laboratories UL 2054,
IEC 60601-1(2005 + A1:2012) Ed 3.1 [when
installed in the CADD®-Solis Model 2110
infusion pump], IEC 60601-1-2 (2014), and IEC
62133 (2002).
The lithium-ion cells comply with Underwriters
Laboratories UL 1642.
IEEE 802.11 a,b,g,n Radio for Network
Communication
The CADD®-Solis Communication Module
intentionally receives and transmits Radio
Frequency (RF) electromagnetic energy for non-
critical communication.
RF transmit / receive, "a" band ....... 5180 MHz to
5320 MHz &
5500 MHz to
5825 MHz
RF transmit / receive, "b", "g" and "n" bands ... 2412
MHz to
2472
MHz
Bandwidth of "a", "b", "g" and "n" signals .........20 MHz
Maximum RF transmit power of band "a" .......10mW
Maximum RF transmit power of bands "b", "g" and
"n" ..................................................................................100mW