Users Manual Part 1

5
IMPORTANT: The rechargeable battery, in the
CADD®-Solis Communication Module, must be
internally connected and fully charged within the
pump before placing the system into service.
Review and follow the instructions in this Operator's
Manual to install and charge the battery prior to use.
Read the instructions completely before installing
the CADD®-Solis Communication Module to
the CADD®-Solis Ambulatory Infusion Pump,
Model 2110. Retain these instructions for future
reference.
For detailed instructions, warnings, specications,
and additional information for operating the CADD®-
Solis Ambulatory Infusion Pump, refer to the pump
Operator’s Manual.
For detailed instructions, warnings, specications, and
additional information for using the CADD®-Solis AC
Adapter, refer to the Instructions for Use supplied with
that product.
WARNINGS
To avoid serious personal injury, it is essential to
observe the following warnings:
Do not expose the communication module to
rain, water, or moisture. Shock or other hazard
may result.
Do not immerse the CADD®-Solis
Communication Module rechargeable battery
in liquid. Shock or other hazard may result.
Do not use the CADD®-Solis Communication
Module rechargeable battery with any device
except the CADD®-Solis Communication
Module.
Do not connect any AC adapters directly to
the CAD-Solis Communication Module
rechargeable battery. Directly connecting
any AC adapter could be hazardous and
can damage the rechargeable battery,
communication module, and the pump.
A CADD®-Solis Communication Module
rechargeable battery that has reached the
end of its useful life must be replaced with
another CADD®-Solis Communication Module
rechargeable battery. Using rechargeable
batteries from other manufacturers could
result in re or explosion.
Do not allow the electronic contacts of the
communication module to simultaneously
touch a piece of metal (a coin or paper clip, for
instance).
Do not use the CADD®-Solis Communication
Module if it has been damaged, if internal
electronics become exposed due to wear, or
if the electronic contacts are damaged in any
way. Fire, shock, or other hazard may result.
Do not attempt to crush the CADD®-Solis
Communication Module or the rechargeable
battery. Fire, shock, or other hazard may
result.
Do not destroy the CADD®-Solis
Communication Module rechargeable battery
by incineration. It may explode.
Do not use the communication module in
the CADD®-Solis Infusion Pump pouch or
any other pouch or bag. The communication
module could overheat and power o utilizing
the thermal protection circuit.
There are potential health hazards associated
with improper disposal of batteries and other
electronics. Dispose of used batteries and
other used accessories in an environmentally
safe manner, and according to any regulations
that may apply.