Users Manual Part 1

7
The CADD®-Solis Network Setup Utility software
(available separately) is used to congure the wireless
network settings for the specic pump when a
CADD®-Solis Communication Module is installed. For
information about using this software, refer to the
installation instructions provided with the software.
NOTE: Version 1.1 of the Network Setup utility must
be used.
WARNING: A CADD®-Solis Communication Module
rechargeable battery that has reached the end of its
useful life must be replaced with another CADD®-
Solis rechargeable battery. Do not use replacement
batteries from other manufacturers as that could
result in re or explosion. For more information, see
the pump Operator's Manual, Rechargeable battery
reached end of use alarm.
WARNING: Do not connect any AC adapters directly
to the CADD®-Solis Communication Module
rechargeable battery. Directly connecting any AC
adapter could be hazardous and can damage the
rechargeable battery, communication module, and
the pump.
Installing the CADD®-Solis Communication
Module
CAUTION: The installation and conguration of the
CADD®-Solis Communication Module in a pump
should only be performed by service personnel.
CAUTION:
To avoid damage to the communication
module, power o the pump and disconnect all
cables connected to the pump prior to installing
the communication module to the pump.
CAUTION: Because the initial installation of the
communication module requires opening the
module housing, this installation should only be
performed at a workstation with electrostatic
discharge (ESD) controls, including a grounded mat
and wrist strap.
1. Make sure the CADD®-Solis pump is powered o
and there are no cables attached to the pump.
2. Using your ngers, the pump key, or a coin, turn
the knob on the battery door counter-clockwise
and open the battery door.
3. Remove any batteries that may be in the pump.
Also, ensure that there is no debris in the battery
compartment.
4. Pull the battery door
rearward, as shown,
to remove it from the
pump. The battery
door is not used with
the CADD®-Solis
Communication
Module. Keep the
battery door in an
accessible location
to allow the pump to
be used later with AA
batteries or the CADD®-