5.3 Measure Info
Measure info
Indicates the average duration. Click on the indicator to modify the Leq Averaging
Time value.
Leq: Elapsed time since the beginning of the current averaging period.
Dataset: Elapsed time since the beginning of the current DataSet.
Since Start: Elapsed time since the start of the measurement.
When using the AutoStore Setup in Multiple Mode, a new average will start at
the end of each Leq period. Those average data are stored in a DataSet, which
may be referred as a data file. Also, a measurement is sometimes partitioned into
several DataSets (usually one day each) to avoid very large files and to facilitate
the post analysis.
Audio Recording State
Transparent background: Input active, but audio recording disabled.
Green background: Audio recording enabled, but currently not recording.
Red background: Audio recording in progress on this input.
Over/Under Range
Transparent background: Input level is between the dynamic range limits.
Blue background (low level): Input level is below the dynamic range lower limit.
Red background (overload): Input level is above the dynamic range upper limit.
Red outline (overload memory): Input level above the dynamic range upper limit
during the active Leq average period.
Photo Info: During the acquisition, it shows the number of photo events.
Photo Viewer: After the acquisition, the button gives access to the Photo Viewer
interface.
The FTP queue indicates the number of elements that are still to be transferred
to the FTP server.
The Output Generation indicator turns on when at least one of the generators is
activated.
The ChOut and PC are also displayed to indicate what generator is activated.
When the cursor passes over the indicator, the following
control appears. This control allows the user to manually
control both generators independently.