Manual

64
1/3 Octave Filter Linearity
The linearity of the 1/3-octave filter has been measured for both ranges (low and high). The
experimentation is done with an adaptor (ADP092) and an electric signal. The results in dB are for an
input sensitivity of 50 mV/Pa. The maximum and the minimum linear levels are measured for each
1/3 octave band along with the noise floor.
Filter Linearity (Low Range)
Frequency (Hz)
Saturation Level
(dB)
Minimum Linear
Level (dB)
Linear Dynamic
Range (dB)
Noise Floor (dB)
120.5 39.5 81.0
1.3
120.5 34.2 86.3
0.7
120.5 33.2 87.3
2.0
120.5 30.6 89.9
1.2
120.5 30.0 90.5
1.1
120.5 28.1 92.4
3.2
120.5 27.8 92.7
0.4
120.5 27.4 93.1
-0.4
120.5 27.2 93.3
1.4
120.5 27.0 93.5
0.0
120.5 26.7 93.8
0.4
120.5 23.4 97.1
0.7
120.5 24.1 96.4
1.3
120.5 23.5 97.0
1.8
120.5 23.5 97.0
2.3
120.5 24.0 96.5
3.2
120.5 24.1 96.4
3.4
120.5 24.1 96.4
4.3
120.5 24.5 96.0
5.2
120.5 24.5 96.0
6.1
120.5 24.2 96.3
7.1
120.5 24.5 96.0
8.2
120.5 24.6 95.9
9.2
120.5 24.7 95.8
10.1
120.5 25.1 95.4
11.3
120.5 25.2 95.3
12.6
120.5 25.8 94.7
14.0