3 Operation Test
Equipment : Gateway(S-Gate), R/C, Jig for power supply
Target DUT : EPD TAG
Figure 3. Operation Test
Test Scenario
Install Tag on Jig
Ready to Gateway(S-Gate) and R/C
Tag start to communicate with Gateway
Tag display the image received from Gateway and enter to sleep mode.
(Display pattern : MAC
SCAN Icon
White)
wake up Tag with R/C
Check Tag (Display pattern : SCAN Icon
White)
Get off Tag from Jig.
Criterion
(Below both two conditions should be all satisfied)
Good : You can see a White display.
Compare MAC address with that of label attached to Side or Back plane.
* gateway lost / not found : You can see a diamond(
) in the upper right corner.