Installation Instructions
33
46900V001_Rev-D_EN
8.  Electrical connection
8.1  Connection to a power socket
A power socket is required for the electrical connection of 
the operator.
A power socket must be installed by trained electricians 
only. The power socket must be protected by a fuse. 
Local and national regulations must be observed 
(e.g. VDE).
People under the influence of drugs, alcohol, or 
medications that can influence their ability to react may 
not work on the operator.
Observe in particular the following safety instructions for 
this chapter.
   DANGER
Danger due to electric current!
Contact with live parts may result 
in electric current owing through 
the body. Electric shock, burns or death 
will result.
 ► All work on electrical components 
may only be carried out by a trained 
electrician.
 ► Before inserting the mains power plug 
forthersttime,ensurethatthevoltage
of the power source matches the 
voltage listed on the operator type plate.
 ► Do not connect the power supply until 
installation is complete.
 ► Disconnect the mains plug before 
working on the operator.
 ► If an accumulator is connected, 
disconnect it from the control unit.
 ► Check that the operator is not live.
 ► Secure the operator against being 
switched back on.
NOTE
Do not connect the ceiling control unit 
to the power supply until the installation 
is complete to prevent damage to the 
operator.
< 1100 mm
Fig. Distance of ceiling control unit to power socket
Note that the distance between the ceiling control unit 
and the power socket must not exceed 1.1 m.
INFORMATION
The power socket must be installed as 
follows:
•  within easy reach of the ceiling control 
unit power cable
•  easily visible and clear of obstacles
INFORMATION
The power cable is approx. 1.2 m long. 
INFORMATION
The mains supply line that has been 
provided may not be shortened or 
extended.
All devices to be connected externally 
must have a safe isolation of the contacts 
from the mains voltage supply according 
to IEC 60364-4-41.
Wiring for external devices must be installed 
in accordance with IEC 60364-4-41.
All electrical wiring must be rmly secured 
to prevent displacement.