User manual

7 Equipment
103
Immunity to series of fast transi-
ents/bursts
IEC 61000-4-4
Amplitude 2 kV, 5 kHz
Immunity to surges
IEC 61000-4-5
Amplitude 2 kV (L-L), 4 kV (L-PE)
Emission of radiated RF disturbances
IEC 61000-6-3
30…230 MHz, 30 dB(V/m) at 10 m
230…1000 MHz, 37 dB(V/m) at 10 m
Emission of conducted disturbances
IEC 61000-6-3
Levels for a quasi-peak detector:
0.15 kHz…0.5 MHz: 66 dBV…56 dBV
0.5 MHz…5 MHz: 56 dBV
5 MHz…30 MHz: 60 dBV
EN 55022 Compliance statement:
PQM-703 is a class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.
6.14 Standards
Standards
Measurement methods
IEC 61000-4-30 Class A
Measurement accuracy
IEC 61000-4-30 Class A
Power quality
EN 50160
Flicker
IEC 61000-4-15
Harmonics
IEC 61000-4-7
Quality standard
design, construction and manufacturing are ISO 9001 compliant
7 Equipment
7.1 Standard equipment
The standard set of equipment supplied by the manufacturer includes:
Permanently fixed cables (2.2 m) with banana plugs (7 pcs)
K01 crocodile clip, black (3 pcs) WAKROBL20K01,
K02 crocodile clip, blue WAKROBU20K02,
crocodile clip, yellow K02 WAKROYE20K02,
red crocodile clip (2 pcs) WAKRORE20K02,
phase splitter AC-16 WAADAAC16,
mains plug with banana inputs (L1 and N) for connecting the analyzer to a socket in order to
charge the battery and/or carry out data transmission from a PC - WAADAAZ1,
software for data reading and analyzing "Sonel Analysis 2",
USB cable WAPRZUSB,
USB wireless module OR-1 WAADAUSBOR1,
a case for the analyzer and standard accessories WAWALXL2,
band for mounting the device on a pole (2 pcs) WAPOZOPAKPL,
bracket for mounting the analyzer on DIN (ISO) rail with stabilizing connectors WAPOZUCH3,
connectors for bands used for mounting the device on a pole (2 pcs) WAPOZUCH4,
voltage adapter with thread M4/M6 (5 pcs) used to connect voltage wires to rail connectors in
switchboards - WAADAM4M6,
magnetic adapter (4 pcs); used to connect voltage wires to overvoltage switches type S and
circuit breakers and RCDs in switchboards - WAADAUMAGKPL,
built-in Li-Ion rechargeable battery,