User Manual
IMPORTANT SAFEGUARDS 
READ BEFORE INSTALLING OR 
USING THE EQUIPMENT 
Your Ultrasonic Processor has been designed with safety in mind. However, no design 
can completely protect against improper usage, which may result in bodily injury and/or 
property damage. For your protection and equipment safeguard, observe the following 
warnings at all times, read the operating instructions carefully before operating the 
equipment, and retain this instruction manual for future reference. If the ultrasonic 
Processor is used in a manner contrary to that specified in this instruction manual, the 
protection features designed into the unit may be impaired. 
•  When mounting the probe, always clamp the converter housing. Never clamp the 
probe. 
•  Make sure the Ultrasonic Processor is properly grounded via a 3-prong outlet. 
•  High voltage is present in the power supply. Do not remove the cover. Refer all 
servicing to qualified service personnel. 
•  To avoid electric shock, disconnect the electrical power cord before removing the 
cover prior to servicing. 
•  Never operate the power supply unless it is connected to the converter. 
•  Never secure anything to the probe, except at the nodal point (point of no activity). 
•  Never touch a vibrating probe. 
•  Never allow a microtip or extender to vibrate in air for more than 10 seconds. 
•  When using a microtip, always keep the amplitude below 40%. 
•  Never operate a probe with threaded end without a tip, extender or microtip. 
•  Air-cool the converter when sample temperature exceeds 100° C, and when working 
at high intensity for more than 30 minutes. 
•  It is recommended that a sound abating enclosure or ear protection be used when 
operating the Ultrasonic Processor. 
WARNING or CAUTION 
Where you see this alert symbol and WARNING or 
CAUTION heading, strictly follow the warning instructions 
to avoid personal injury or equipment failure. 