Guidelines and Manufacturer's Statement--Electromagnetic Immunity
The SoundBite Hearing System is expected to be used in the following specified electromagnetic environment, and the purchaser or user
shall ensure that it is used in this electromagnetic environment:
Immunity test
Basic EMC standard or test
method
Home healthcare facility environment
Electrostatic discharge
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV, +/- 8 kV, +/- 15 kV air
Radiated RF EM fields
IEC 61000-4-3
10V/m
80MHz-2.7GHz
80%AM at 1kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
See the RF wireless communication equipment table in
"Recommended minimum separation distances".
Rated power frequency magnetic
fields
IEC 61000-4-8
30A/m; 50 Hz or 60Hz
Electric fast transients bursts
IEC 61000-4-4
±2kV
100kHz repetition frequency
Surges
IEC 61000-4-5
Line to line: ±0.5kV, ±1kV
Conducted disturbances induced
by RF fields
IEC 61000-4-6
3 V in 0.15 MHz - 80 MHz
6 V in ISM and/or amateur radio bands between 0.15 MHz and 80
MHz
80 % AM at 1kHz or 2 Hz
Voltage dips
IEC 61000-4-11
0% U
T
: 0.5 cycle
at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°