If the system has generated the interference (confirmed by powering on and
off the system), the qualified service personnel should solve the problem by
following the steps as below
- Change the location of the affected system.
- Place SonoMe further away from the affected system.
- Supply power to this system in other ways other than the way used currently
- Contact Bionet’s Customer Service Department ASAP
Warning
Use of this equipment adjacent to or stacked with other equipment may result in
improper operation and should be avoided. If such use is required, this and other
equipment should be observed to ensure normal operation.
Warning
The use of accessories, converters and cables not specified or provided by the
manufacturer may increase electromagnetic emissions or decrease electromagnetic
immunity of this equipment and may cause malfunction.
Caution
The emission characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 Class A). For use in a residential environment (which
normally requires CISPR 11 Class B), this equipment may not provide adequate
protection for radio frequency communications services. Users may need to take
mitigation measures such as relocating or reorienting equipment.
Warning
This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions: