User's Guide

28
Appendix A Specifications
Complied
Standards
EN 60601-1 (IEC 60601-1), Medical electrical equipment Part 1: General
requirements for basic safety and essential performance, Class I, BF, continuous
operation EN 60601-2-37:2008 (IEC 60601-2-37:2007), Medical Electrical
Equipment Part 2-37: Particular Requirements for the Basic Safety and Essential
Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment EN
60601-1-2:2007 (IEC 60601-1-2: 2007), Class A
Safety
Types
Type of protection
against electric
shock
Internally powered
Degree of protection
against electric
shock
Type-BF applied part
Operation mode
Continuous working
Installation and
operation type
Portable Equipment
Degrees of
protection against
harmful liquid
IPX5
Degree of safety of
application
The equipment is not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen or nitrous
oxide.
Environm
ental
Requirem
ent
Operations
Storage and Transportation
Relative Humidity
25% to 80%,
non-condens
ing
25% to 93%, non-condensing
Ambient
Temperature
5°C to
+40°C
-20°C to +55°C
Atmospheric
Pressure
700hPa to
1060hPa
700hPa to 1060hPa
Max. Altitude
3000m
3000m